Static pressure
(zero errors can be calibrated out at line pressure)
per 0.5 MPa, 5 bar or 72.5 psi (sensor A)
per 2 MPa, 20 bar or 290 psi (sensor B)
per 3.5 MPa, 35 bar or 500 psi (sensor E)
per 7 MPa, 70 bar or 1015 psi (sensor F to S)
Model 266DSH standard static
• zero error: ±0.05 % of URL for sensor F to S
±0.08 % of URL for sensor A, B and E
• span error: ±0.08 % of reading.
Model 266DSH flange mounted
per 2 MPa, 20 bar or 290 psi
• zero error: ±0.05 % of URL for sensor F to P
±0.08 % of URL for sensor E
span error: ±0.08 % of reading.
Mounting position
Rotations in the plane of the diaphragm have a negligible
effect. A tilt from the vertical of up to 90° causes a zero
point shift of up to 0.5 kPa (5 mbar, 2 inH2O), which can be
corrected by making an appropriate zero position
adjustment. There is no effect on the measuring span.
In applications where a dynamic inclined installation is
required (e.g. naval applications with maximum vessel
inclination of +/-22.5°), the maximum error can be up to 3
mbar for 266DSH
Such error might be resulting in a deviation from the
standard accuracy with percentual magnitude depending
on the range of the selected device.
In case of non-dynamic inclined installations, this effect
can be eliminated by performing a zero calibration in the field.
Please contact ABB to assess more in details on the installation effects