Standard features
Modular design includes
• Modular software – application based plug in software modules
• Manifold module (internal tubing now replaced
by manifold)
• Analysis section contains stream selection solenoids,
pressure regulation, 32 bit digital detector electronics and
a dual-train chromatograph in a single, replaceable module
• 32 bit digital, low-power controlling electronics, using
Windows CE® (internal to GC unit)
• Microsoft® Windows® based man-machine interface software (PCCU 32)
• Lithium battery-backed RAM
• Two remote serial digital communications ports; one localport
• Comprehensive diagnostics and wizards available to users
• Three-level security for user access control
• Audit-quality historical data; date and time stamped
Auto-start cycle automatically
• Stabilizes oven temperature
• Confirms modules’ functions
• Sets valve operation times
• Locates natural gas components
• Sets carrier gas pressures
• Gates natural gas components
• Validates/calibrates
• Returns to pipeline streams
• Operational alarms and events, time, and data stamped in a log.
• Memory: up to 2.052 GB
• Detectors - constant temperature, glass encapsulated
thermister beads for rugged service and long life. Will not
burn out on loss of carrier.
• Dual 10 port valves have no moving metal parts.
• Low utility usage - low-power, low-carrier, no instrument air required
• On demand or scheduled automatic calibration and diagnostics
Calibration options
• Exponential
• Linear
• Multi-point (6 points)
wechat/whatsapp: