Standard features
Modular design includes
• Modular software – application based plug in software modules
• Manifold module (internal tubing now replaced
by manifold)
• Analysis section contains stream selection solenoids,
pressure regulation, 32 bit digital detector electronics and
a dual-train chromatograph in a single, replaceable module
• 32 bit digital, low-power controlling electronics, using
Windows CE® (internal to GC unit)
• Microsoft® Windows® based man-machine interface software (PCCU 32)
• Lithium battery-backed RAM
• Two remote serial digital communications ports; one localport
• Comprehensive diagnostics and wizards available to users
• Three-level security for user access control
• Audit-quality historical data; date and time stamped
Auto-start cycle automatically
• Stabilizes oven temperature
• Confirms modules’ functions
• Sets valve operation times
• Locates natural gas components
• Sets carrier gas pressures
• Gates natural gas components
• Validates/calibrates
• Returns to pipeline streams
• Operational alarms and events, time, and data stamped in a log.
• Memory: up to 2.052 GB
• Detectors - constant temperature, glass encapsulated
thermister beads for rugged service and long life. Will not
burn out on loss of carrier.
• Dual 10 port valves have no moving metal parts.
• Low utility usage - low-power, low-carrier, no instrument air required
• On demand or scheduled automatic calibration and diagnostics
Calibration options
• Exponential
• Linear
• Multi-point (6 points)
—
Standard options
• Sample Conditioning System
• On board digital ¼ VGA display with multiple screen access
• USB (Host and Client) and Ethernet ports
• Calculation packages (eg, V.O.S., HCDP, Etc.)
• SD memory cards (up to 2 GB)
• Feed-through heater
• C7+, C9+ analysis
• Optional: AI/DI, AO/DO per customer requirements
• Optional: 24 hour temperature testing
—
Maintenance
The NGC8206 was designed from the ground up to be
maintained by personnel with little or no prior knowledge
of gas chromatography.
Both hardware and software are designed to provide low
maintenance through easily replaceable electro-mechanical
modules such as:
• NGC termination panel
• NGC analytical module, an easily removable chromatograph
subassembly, containing:
- NGC manifold
- Stream selector solenoid valves
- GC module
- Dual electronic carrier pressure regulator valves
- Chromatograph pilot valve
- NGC analytical processor
• NGC feed-through (heater optional)
• Sample conditioning module (SCM)
• NGC controller, (32 bit processor)
• NGC ¼ VGA display (optional)
Other maintenance support features
• Intuitive Local Operator Interface (PCCU 32) running
Microsoft® Windows® environment
• Diagnostic software and wizards for maintenance
• Diagnostic file output system for e-mail support
• Digitized detector output (chromatogram) to PCCU 32
• Remote or local operation of PCCU 32
• Quick Start guide, Start up manual and Start up video
wechat/whatsapp:
+86-181-4410-0983
Email: kongjiangauto@163.com