Introduction
Specialty analysis for separation of wide boiling range
compounds. The PGC5000 Temperature Programmed (TP)
Gas Chromatograph uses component separation to analyze
gas or liquid samples. The analyzer operates unattended,
automatically sampling and analyzing process streams.
—
Description
An isothermal oven contains the temperature programmed
oven assembly with a detector, either a Flame Ionization
Detector (FID) or Thermal Conductivity Detector (TCD). The
detector is mounted on top of the temperature programmed
oven, and it is integrated with the TP oven in order to provide
an optimal analytical flow path and eliminate potential cold
spots. The sample valves, either a Liquid Sample Valve (LSV)
or a Continuous Performance (CP) valve, can also be installed
in the isothermal oven compartment. An optional air
cleanup / methanizer unit may be used. Special columns
(capillary or packed type) are installed inside the temperature
programmed oven. The columns contain special packing
material which separates the compounds to be analyzed so
that they enter the detector in a predictable sequence.
—
Physical
Environmental (enclosure)
Protected from weather: IP 54. (NEMA Type 13) equivalent
Ambient temperature range
0 to +50 °C (32 to 122 °F)
Humidity
95% relative humidity, non-condensing
Dimensions (W x D x H)
496 x 340 x 1175 mm
(19.5 x 13.4 x 46.3 in.)
Weight
73 kg (160 lb) (minimum)
Mounting
Wall: 32 mm (1.3 in.) from wall with brackets
Floor: Optional wheeled dolly
EMI/RFI considerations
Conforms to Class A industrial environment
Electrical entries
Left side
Pneumatic entries
Right side
Sample entries
Liquid
Right side, 1 each model 791 LSV
Gas
Bottom and/or right side
Vents
Bottom and/or right side