Safety area classification
CSA / NRTL: Class I, Division 1; gas groups B, C, D with type Y-purge
Class I, Division 2; gas groups B, C, D
temperature code T4 – T2
ATEX / IEC / CN / KO: Zone 1: CE 0344; II2G, Ex de py IIB+H2 T4 – T2
Zone 2: CE; II3G Ex de nA nL IIB+H2 T4 – T2
Ex de px IIB+H2 T4 – T2 (optional)
With X-purge power interlock
Purge wait time: 18 minutes (Class I, Division 1 / Zone 1 area)
Power (hot, neutral, ground)
Voltage: 100 – 240 VAC
Frequency: 50-60 Hz
Power consumption: 1,200 Watts startup, 900 Watts steady-state operation
Typical, varies with installed options.
Instrument air
Supply connection: 3/8 inch tube, minimum
Supply pressure: 551.6 kPa (80 psig)
Quality: Instrument grade: Clean, oil free and -34° C, (-30° F) dewpoint
Flow rates: Steady state purge: 127-147 L/min (4.5-5.2 ft3/min) at 20° C, Y-purge types
Analytical detectors
Standard detectors: Single and multiport thermal conductivity, flame ionization, flame photometric
Third party detectors: Consult factory for availability