Description
The truck is moved as close to the sample point as is practical.
A flexible sample line is then uncoiled and attached to the probe
in the flowing natural gas pipeline. If the dew point of the gas or
the weather demands it, this flexible tubing must be heated. A
sample of the gas is extracted from the probe in the line, purged
through the flexible line and then transported to the analyzer.
There it is processed for particle removal and phase integrity,
injected onto the chromatographic columns where component
separation occurs. These components are then quantified by
peak integration and concentration is calculated. The other data
of interest is then calculated from these concentrations.
The Portable NGC8209 analyzes each sample utilizing
established chromatographic techniques. The resulting
information consists of mole percent values for the following:
– Air (Contains N2, Ar, CO and O2 but not CO2), C1, CO2, C2,
C3, IC4, NC4, NeoC5, IC5, NC5, C6’s, C7’s, C8’s, C9+
The NGC8209 measures a C9+ (back flushed) peak. Users may
input the results of a comprehensive lab analysis that reflects
the split or ratio of C9 through C12 components. This ratio can
be used in subsequent analyses and energy calculations.
Calculated values include
– Relative density (specific gravity)
– Btu/CV (heating) value
– GPM (gallons of liquid per thousand cubic feet)
– Wobbe index
Historical data
The NGC8209 is designed to retain historical data. This data
can be used for custody transfer needs, verify transmitter
operation over time, and provide a limited data backup for
communication link reliability. In addition, various CSV file
outputs are available to download the analysis data into
spreadsheet based company specific data sheets, and
outputs to 3rd party gas accounting packages.
The default¹ memory configuration provides the most recent
480 analysis cycles containing:
– Normalized components
– Un-normalized components
– Ideal Btu/CV
– Real Btu (wet and dry) / CV (superior and inferior)
– Relative density (specific gravity)
– Density
– GPM
– Wobbe index
– Alarms
Stream averages for the (default¹) 840 last hours, 35 last
days and the most recent last month analyses. Operational
Parameters for the (default¹) last 480 cycles (Diagnostics Report):
– Selected peak times
– Selected peak areas
– Ideal Btu/CV
– Carrier regulator pressure
– Oven temperature
– Ambient temperature
– Sample pressure
– Detector noise values
– Detector balance values
Audit logs (Default¹)
– Last 480 alarms
– Last 480 events
Data retained by the NGC8209 can be collected via a remote
communication link or by the laptop PC local operator
interface PCCU 32.
Equipment
The equipment used for this analysis consists of the standard
NGC8209 mounted in a black case. It is recommended that
the unit be left in an operational mode while transporting.
All the valves, gauges, inlet ports, and PCCU32 (the laptop
driven man-machine interface) connectors are bulkhead
mounted on panel to provide the technician with ready access.