Historical data
The NGC8106 is designed to retain historical data. This data
can be used for audit trails, maintenance, and troubleshooting
to verify transmitter operation over time and provide a limited
data backup for communication link reliability.
The user is allowed to configure the period of the data
retained by the NGC8106 via the Operator Interface. The
default1 memory configuration provides the most recent 480
analysis cycles containing:
Stream averages for the (default1) 840 last hours, 35 last
days and the most recent last month analyses. Operational
parameters for the (default1) last 480 cycles (Diagnostics
Report): Audit logs (default1); Data retained by the NGC8106
can be collected via a remote communication link or by the laptop
PC local operator interface PCCU.
Available accessories
– 120/240 Vac to 12Vdc GP and exp power supplies
– Pole or pipeline mounting kits
– Cold weather enclosure (also available in pipe mount
configuration)
– Regulators (carrier and calibration blend)
– SD memory card
– Export crating
– Tool kit
– Various maintenance kits
Installation dimensions
Specifications
– Designed for gas streams with the following characteristics:
– Clean is defined as having no particles larger than 1
micron and no more than 1 milligram of solids per cubic
meter of gas
– Dry is defined as no more than 7 psi of water per million
cubic feet of gas
– Dry is also defined as having less than 0.1 ppm of liquid
at the coldest ambient condition expected at the coldest
point in the system (the liquid can be water, oil, synthetic
lubricant, glycol, condensed sample, or any other nonvapor contaminate)
– Stable gas is defined as a vapor containing less than 0.1 ppm
of liquid when the vapor is cooled to 10°C below the coldest
ambient temperature possible at any point in the system
– Single stream capability. Manual calibration required