Flow Rate Control using Flow Restrictors
The flow rates of test gas and reference air are important to the
accurate operation of Zirconia-based oxygen analyzers and,
traditionally, control has been achieved using ancillary
flowmeters and flow control valves.
The Endura AZ20 offers a new approach with the inclusion of
optional flow restrictors. These are fitted within the probe to
control the flow rate of the reference air and test gasses,
removing the need for additional flowmeters fitted with flow control valves.
This innovative approach enables the gasses to be controlled by
pressure only. 1 Bar (15 psi) is applied directly to the probe’s gas
ports and the correct flow rate is then set by the flow restrictors.
Unique Integrated Automatic Calibration
The all new Endura AZ20 automatic calibration system
eliminates the need for the expensive ancillary equipment
required for automatic calibration on traditional flue gas oxygen
analyzer systems. ABB’s fully integrated, automatic calibration
feature controls the test gas sequence and detects test gas
availability, eliminating incorrect calibrations due to loss of test gas.