Measurement principle
Non-dispersive infrared absorption photometers in the λ = 2.5-8 µm wavelength range can
measure up to 4 compositions using 1 or 2 beam paths and 1 or 2 receivers per beam path in one gas path or two independent gas paths.
Measure up to 4 components in one gas path or two independent gas paths with 1 or 2 receivers per beam path.
Impact Effects
Flow rate effects
Flow rates in the range of 20-100 liters/hour: within the detection range
Correlated gas effects/cross sensitivity
Analyzer configuration must be aware of sample gas composition.
Selective measures to reduce associated gas effects (optional): Installation of interference filters or filter units,
internal electronic cross-sensitivity correction of one sample component by other sample components measured by the gas analyzer.
Temperature effects
Ambient temperature within the permissible range
- At zero point: ≤ 2 % of range per 10 °C - At sensitivity without thermostat:
≤ 3 % of the measured value per 10 °C - Sensitivity with thermostat (optional):
Sensitivity with thermostat (optional): ≤ 2 % of measured value per 10 °C
Thermostat temperature = 55 °C
Barometric influences - At zero point: No influence - Sensitivity during pressure correction by means of integral pressure sensor
Per 1 % barometric pressure change: ≤ 0.2 % of measured value
Barometric pressure variation
If a hose is used for the internal gas line, the pressure sensor is located in the sampling gas line.
If a hose is used for the internal gas line, the pressure sensor is routed externally through the hose.
Operating range of the pressure sensor: pabs = 600-1250 hPa
Power supply influences
Voltage and frequency within the permissible range: no effect