CLS200 Specifications
Operator Interface
• 32-character vacuum fluorescent display
• 8-key keypad to access guided menus and prompts,
enter passkey sequence, set values, switch between
single channel and multiple channel displays
• Controller configuration can be loaded through the
standard serial port
Analog Inputs
• CLS204 4 differential
• CLS208 8 differential
• CLS216 16 single-ended
Noise Rejection
• 120dB at 60Hz
Temperature Coefficient
• 40ppm/°C
Sensors/Inputs
• Thermocouples: user-selectable type, direct connection,
linearization, reference junction compensation, reversed
and shorted thermocouple detection and upscale break
protection with output averaging
• RTD: (CLS204 and CLS208 only) 2- or 3-wire, platinum,
100Ω @ 0°C, DIN 0.003850Ω/Ω/°C curve, user-selectable
range. Two user-selectable ranges offer different
resolutions. Requires scaling resistors. See special/linear
inputs in ordering information.
• Linear: current and voltage signals from linear transmitters
• Pulse input
Input Range and Accuracy
Sensor Range (°F) Range (°C) Accuracy
Type B 150 to 3200°F 66 to 1760°C ±4.0°C
Type E -328 to 1448°F -200 to 787°C ±1.0°C
Type J -350 to 1400°F -212 to 760°C ±1.2°C
Type K -450 to 2500°F -268 to 1371°C ±1.3°C
Type R 0 to 3210°F -18 to 1766°C ±2.8°C
Type S 0 to 3200°F -18 to 1760°C ±2.8°C
Type T -450 to 750°F -268 to 399°C ±1.6°C
RTDs available on CLS204 and CLS208 only.
Sensor Range (°F) Range (°C) Accuracy
RTD1 -148 to 527˚F -100 to 275˚C ±1.1˚C
RTD2 -184 to 1544˚F -120 to 840˚C ±1.6˚C
Note: Accuracy @ 25˚C ambient. Valid for 10 to 100 percent
of span except Type B, which is specified for 800˚F to
3200˚F. RTD is for 100 percent of span.
Linear Voltage and Current Inputs
Requires scaling resistors. See special inputs in ordering
information.
0-20mA(dc)/4-20mA(dc)
0-5VDC
0-10VDC
Other ranges available. Contact factory.
Pulse Input
• One TTL-level square wave input up to 2kHz
Input Sampling Rate at 60Hz
Each channel has the following scans per second:
• CLS204: 6 samples per second (update time: 0.167 sec.)
• CLS208: 3 samples per second, (update time: 0.333 sec.)
• CLS216: 1.5 samples per second, (update time: 0.667 sec.)
Internal Measurement Resolution
• 0.006 percent, greater than 14 bits
Calibration
• Automatic zero and full scale
Digital Inputs
• TTL-level used for selecting recipes or jobs, or R/S triggers
• 8 inputs and 1 pulse input with 50-pin terminal board
option
• 2 inputs and pulse input or 3 inputs with 18-pin terminal
block option
Digital Outputs
• 34 digital outputs are available with 50-pin terminal
board option
• 10 control outputs with 18-pin terminal block option
• 1 or 2 control outputs are user assigned for each loop
• Each control output can be configured for on-off time
proportioning, or distributed zero crossing
• Outputs sink up to 60mA each at 5VDC
• 350mA at 5VDC available from on-board supply
Alarm Outputs
• Independent process and deviation alarms for each
channel
• Alarms can operate any output not used for control
• User-programmable deadband, delay and startup
suppression
• Global alarm output activates when any alarm occurs
• Watchdog output indicates controller is functioning
correctly
Serial Interface
• EIA/TIA-232 or EIA/TIA-485
Baud Rate
• 2400. 9600 or 19200. user-selectable
Communications Protocols
• Form of ANSI X3.28-1976. (D1. F1) compatible with
Allen-Bradley PLC/2
• Modbus® RTU
Line Voltage/Power
• 15 to 24VDC ± 3VDC @ 1A (max.),
300mA (no load)
Agency Approvals
• UL®, C-UL® listed: UL® 61010-1 safety requirements for
measurement, control and laboratory equipment
• CE Mark: See Declaration of Conformity for details