ATEX Directive 2014/34/EU, annex II
The PGC5000 analyzer “must be so designed and
constructed as to prevent ignition sources arising,
even in the event of frequently occurring
disturbances or equipment operating faults, which
normally have to be taken into account”. The design
of the analyzer is capable of functioning within the
established operational parameters and of ensuring
a high level of protection necessary for the intended
classified Ex zone 1 area.
The notified body LCIE examined the internal components of the PGC5000
and approved the analyzer as Ex ‘py’ for a zone 1
area:
• The EC-type examination of the PGC5000 series
analyzer and the IEC Ex test report first evaluated
the internal components to Part 15 of the
harmonized conformity standard IEC60079 (Ex ‘n’)
as either being non-sparking, energy limited or
temperature limited to below the specified
temperature class (ref: certificates EC-type LCIE
09ATEX 1017X and IECEx LCI09.0010X).
• The EC-type examination of the PGC5000 series
analyzer and the IEC Ex test report then evaluated
the pressurization protection of the enclosure to
part 2 of the harmonized conformity standard
IEC60079 (Ex ‘p’) as being suitable for zone 1 area
without electrical power interlock to the
pressurization system (ref: certificates EU-type
LCIE 09ATEX 3006X and IECEx LCI09.0010X).